Digox 601 DAC
In order to ensure a short start-up phase, it has to be differentiated whether an air-inleakage or a coolant leakage exists. Therefore, it is necessary to remove the carbonic acid from the sample. The conductivity after the CO2-degassing is now measured (degassed conductivity). If the conductibility value measured after degassing falls under the value of 0.2µS cm-1 then merely H+ and OH- according to the ion product of the water are still present as well as slip-induced anions as charge carriers. Thus, a cooling water hardness change can be ruled out and the start-up phase can be significantly shortened.
With the Digox 601 dac, you have a universal measuring instrument at your disposal. In the compact design, the specific conductivity and the cation conductivity are measured and the pH-value is calculated - the "degassed conductivity" is displaye
Liitteet:
Digox 601 dac-E-2011-01.pdf
Ominaisuudet:
Advantages:
- Removes selectively CO2
(no unintentional removal of organic acids) - Measurement of degassed conductivity at ambient temperature
(no T-compensation for high temperatures necessary)
Tuote: Digox_601_dac